Annual Linac QA

Dosimetry:


X-ray flatness change from baseline 1%
X-ray symmetry change from baseline 1%
Electron flatness change from baseline 1%
Electron symmetry change from baseline 1%
SRS arc rotation mode:   (range: 0.5–10 MU/deg)

Monitor units set vs delivered: for SRS/SBRT

1.0 MU or 2% (whichever is greater)

Gantry arc set vs delivered: 1.0° or 2% (whichever is greater)